Ann Lab Med 2019; 39(1): 50-57  https://doi.org/10.3343/alm.2019.39.1.50
Performance Evaluation of the Automated Fluorescent Immunoassay System Rotavirus Assay in Clinical Samples
Jae-Seok Kim, M.D.1, Su-Kyung Lee, M.S.2, Dae-Hyun Ko, M.D.2, Jungwon Hyun, M.D.2, and Hyun Soo Kim, M.D.2
1Department of Laboratory Medicine, Kangdong Sacred Heart Hospital, Hallym University College of Medicine, Seoul, Korea; 2Department of Laboratory Medicine, Hallym University Dongtan Sacred Heart Hospital, Hallym University College of Medicine, Hwaseong, Korea
Corresponding author: Hyun Soo Kim, M.D.
https://orcid.org/0000-0002-7026-6715
Department of Laboratory Medicine, Dongtan Sacred Heart Hospital, Hallym University College of Medicine, 7 Keunjaebong-gil, Hwaseong 18450, Korea
Tel: +82-31-8086-2775
Fax: +82-31-8086-2789
E-mail: hskim0901@empas.com
Received: November 22, 2017; Revised: April 5, 2018; Accepted: August 16, 2018; Published online: January 1, 2019.
© Korean Society for Laboratory Medicine. All rights reserved.

This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
Abstract
Background: The Automated Fluorescent Immunoassay System (AFIAS) rotavirus assay (Boditech Med Inc., Chuncheon, Korea) is a new rapid antigen test for rotavirus detection. We evaluated the performance of this assay for detecting rotaviruses and their specific genotypes in clinical stool samples.
Methods: AFIAS rotavirus assay was performed in 103 rotavirus-positive and 103 rotavirus-negative stool samples (confirmed by both PCR and ELISA), and its results were compared with those of PCR, ELISA, and immunochromatographic assay (ICA). We evaluated diagnostic sensitivity/specificity, the detectability of rotavirus subtypes, lower limit of detection (LLOD), reproducibility, cross-reactivity, and interference of AFIAS rotavirus assay.
Results: Based on PCR and ELISA results, diagnostic sensitivity and specificity of the AFIAS rotavirus assay were both 99.0%. LLOD results showed that the AFIAS assay had sensitivity similar to or greater than ICA and ELISA. High reproducibility was confirmed, and no cross-reactivity or interference was detected. This assay could detect genotypes G1P[8], G2P[4], G3P[8], G4P[6], G4P[8], G8P[4], G8P[8], G9P[4], and G9P[8].
Conclusions: The AFIAS rotavirus assay showed high reproducibility, sensitivity, and specificity as well as excellent agreement with ELISA, PCR, and ICA. It detected the most common as well as unusual genotypes of rotavirus prevalent in Korea. It could be a useful on-site assay for rapid, convenient, and cost-effective detection of rotavirus infection.
Keywords: Rotavirus, AFIAS rotavirus assay, Immunoassay, Performance, Genotype