EDTA Treatment for Overcoming the Prozone Effect and for Predicting C1q Binding in HLA Antibody Testing
Hoon Seok Kim , M.D., Ae-Ran Choi , M.T., Mina Yang , M.D., and Eun-Jee Oh , M.D., Ph.D.
Department of Laboratory Medicine, Seoul St. Mary’s Hospital, College of Medicine, The Catholic University of Korea, Seoul, Korea
Corresponding author: Eun-Jee Oh, M.D., Ph.D.
Department of Laboratory Medicine, Seoul St. Mary’s Hospital, College of Medicine, The Catholic University of Korea, 222 Banpo-daero, Seocho-gu, Seoul 06591, Korea
Tel: +82-2-2258-1641 Fax: +82-2-2258-1719 E-mail: ejoh@catholic.ac.kr
Received: January 4, 2019; Revised: April 1, 2019; Accepted: June 9, 2019; Published online: November 1, 2019.
© Korean Society for Laboratory Medicine. All rights reserved.

This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
Abstract
The Luminex-based single antigen bead (SAB) assay is widely used to detect HLA antibody in transplant recipients. However, one limitation of the SAB assay is the prozone effect, which occurs mostly as a result of complement interference. We investigated the efficacy of EDTA treatment for overcoming the prozone effect and predicting C1q binding of HLA antibody. We subjected 27 non-treated (naïve) and EDTA-treated serum samples from highly sensitized patients to IgG-SAB assays, and we confirmed the prozone effect in 53% and 31% of class I and class II antibody tests, respectively, after EDTA treatment. When we conducted additional assays after dithiothreitol treatment and serum dilution, EDTA was the most efficacious in eliminating the prozone effect. Reducing the prozone effect by EDTA treatment strengthened the correlation between IgG mean fluorescence intensity (MFI) and C1q MFI values (ρ=0.825) as compared with the naïve sera (ρ=0.068). Although C1q positivity was dependent on the concentration of HLA antibody in EDTA-treated sera, the correlations varied individually. Overall, our results confirmed the efficacy of EDTA treatment for overcoming the prozone effect. EDTA treatment showed a positive effect on the correlation between IgG MFI and C1q MFI values.
Keywords: Prozone effect, EDTA, Single antigen bead assay, C1q binding, HLA antibody