Fig. 2.

Download original image

Fig. 2.

Box-and-whisker plot of the differences between each analyzer and HPLC-MS HbA1c results for 15 known Hb variants. The central box represents the interquartile range, and the middle line indicates the median (G11vr, 0.2; G8vr, −0.5; G11st, −1.3; G8st, −1.2). A line extends from the minimum to the maximum value, excluding single outlier in G8st, which is displayed as a separate point for G8st.

Abbreviations: G11vr, HLC-723 G11 Variant Mode (Tosoh Bioscience, Inc., Tokyo, Japan); G8vr, HLC-723 G8 Variant Mode (Tosoh Bioscience); G11st, HLC-723 G11 Standard Mode (Tosoh Bioscience); G8st, HLC-723 G8 Standard Mode (Tosoh Bioscience).

Ann Lab Med 2019;39:237~244 https://doi.org/10.3343/alm.2019.39.3.237