OPEN ACCESS pISSN 2234-3806
eISSN 2234-3814

Table. 3.

Differences between measurement systems after recalibration

Samples Representative values after recalibration (nmol/L) Inter-MS %CV (after recalibration)

Roche Abbott Siemens Beckman Snibe Mindray
Insulin
202111 0.047 0.159 0.122 0.104 0.073 0.063 44.1
202113 0.346 0.315 0.318 0.325 0.336 0.319 3.8
202114 0.367 0.376 0.367 0.373 0.358 0.342 3.6
C-peptide
202111 0.230 0.260 0.310 0.294 0.269 0.289 10.4
202113 2.195 2.204 2.142 2.191 2.287 2.204 2.1
202114 2.867 2.785 2.781 3.113 3.015 2.867 4.6

Abbreviation: MS, measurement systems.

Ann Lab Med 2022;42:428~437 https://doi.org/10.3343/alm.2022.42.4.428

© Ann Lab Med