OPEN ACCESS pISSN 2234-3806
eISSN 2234-3814

Table. 2. Intra- and inter-run precision and trueness
C-peptide Target value (ng/mL) Precision Trueness
Intra-run (N=5) Inter-run (N=20)
Mean (ng/mL) % CV Mean (ng/mL) % CV Mean (ng/mL) % Bias
CRM 6901-C 104±5 ng/mL (1/200 dilution) 0.52 0.50 4.8 0.52 4.5 0.52 0.6
CRM 6901-C 104±5 ng/mL (1/100 dilution) 1.04 1.03 3.6 1.07 4.3 1.05 1.0
CRM 6901-C 104±5 ng/mL (1/10 dilution) 10.40 10.74 4.2 10.66 3.8 10.72 3.1

Abbreviation: CRM, certified reference material.

Ann Lab Med 2025;45:62~69 https://doi.org/10.3343/alm.2024.0072