Prevalence of a Single-Nucleotide Variant of SARS-CoV-2 in Korea and Its Impact on the Diagnostic Sensitivity of the Xpert Xpress SARS-CoV-2 Assay
2022; 42(1): 96-99
Ann Lab Med 2019; 39(6): 566-571
Published online November 1, 2019 https://doi.org/10.3343/alm.2019.39.6.566
Copyright © Korean Society for Laboratory Medicine.
Hyeyoung Lee , M.D.1,2, Ji Hyeong Ryu , B.S.3, Hye-Sun Park , M.S.3, Ki Hyun Park , M.S.3, Hyunjoo Bae , B.S.3, Sojeong Yun , B.S.3, Ae-Ran Choi , B.S.1, Sung-Yeon Cho , M.D.4,5, Chulmin Park , Ph.D.5, Dong-Gun Lee , M.D.4,5, Jihyang Lim , M.D.6, Jehoon Lee , M.D.6, Seungok Lee , M.D.7, Soyoung Shin , M.D.8, Haeil Park , M.D.9, and Eun-Jee Oh , M.D.1
1Department of Laboratory Medicine, Seoul St. Mary’s Hospital, College of Medicine, The Catholic University of Korea Seoul, Korea; 2Department of Laboratory Medicine, Catholic Kwandong University International St. Mary’s Hospital, Incheon, Korea; 3Department of Convergence Medical Science, Graduate School, The Catholic University of Korea, Seoul, Korea; 4Division of Infectious Diseases, Department of Internal Medicine, College of Medicine, The Catholic University of Korea, Seoul, Korea; 5Vaccine Bio Research Institute, College of Medicine, The Catholic University of Korea, Seoul, Korea; 6Department of Laboratory Medicine, Eunpyeong St. Mary’s Hospital, College of Medicine, The Catholic University of Korea Seoul, Korea; 7Department of Laboratory Medicine, Incheon St. Mary’s Hospital, College of Medicine, The Catholic University of Korea Seoul, Korea; 8Department of Laboratory Medicine, Daejeon St. Mary’s Hospital, College of Medicine, The Catholic University of Korea Seoul, Korea; 9Department of Laboratory Medicine, Bucheon St. Mary’s Hospital, College of Medicine, The Catholic University of Korea Seoul, Korea
Correspondence to: Eun-Jee Oh, M.D., Ph.D.
Department of Laboratory Medicine, Seoul St. Mary’s Hospital, College of Medicine, The Catholic University of Korea, 222 Banpo-daero, Seocho-gu, Seoul 06591, Korea
Tel: +82-2-2258-1641 Fax: +82-2-2258-1719 E-mail: ejoh@catholic.ac.kr
This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
ELISAs and rapid diagnostic tests (RDTs) are widely used for diagnosing dengue virus (DENV) infection. Using 138 single blood samples, we compared the ability to detect non-structural (NS)-1 antigen and anti-DENV IgM/IgG antibodies among (1) DENV Detect NS1 ELISA, DENV Detect IgM capture ELISA and DENV Detect IgG ELISA (InBios International, Inc.); (2) Anti-Dengue virus IgM Human ELISA and Anti-Dengue virus IgG Human ELISA (Abcam); (3) Dengue virus NS1 ELISA, Anti-Dengue virus ELISA (IgM) and Anti-Dengue virus ELISA (IgG) (Euroimmun); (4) Asan Easy Test Dengue NS1 Ag 100 and Asan Easy Test Dengue IgG/IgM (Asan Pharm); (5) SD BIOLINE Dengue Duo (Standard Diagnostics); and (6) Ichroma Dengue NS1 and Ichroma Dengue IgG/IgM (Boditech Med). For NS1 antigen detection, InBios and Euroimmun showed higher sensitivities (100%) than the RDTs (42.9–64.3%). All tests demonstrated variable sensitivities for IgM (38.1–90.5%) and IgG (65.7–100.0%). InBios and Boditech Med demonstrated higher sensitivity (95.6% and 88.2%, respectively) than the other tests for combined NS1 antigen and IgM antibody. Five NS1 antigen tests had good agreement (92.8–98.6%) without showing positivity for chikungunya. However, all IgG tests demonstrated potential false-positivity with variable ranges. Clinical laboratories should note performance variations across tests and potential cross-reactivity.
Keywords: Dengue virus, Diagnosis, Rapid diagnostic tests, ELISA, Performance
Dengue virus (DENV) infection is a mosquito-borne disease that constitutes one of the major public health problems in subtropical and tropical areas [1,2]. As many patients either have no symptoms or present with nonspecific fever requiring differential diagnosis, laboratory confirmation using a rapid, accurate, and relatively low-cost diagnostic test is especially important [3]. Laboratory diagnosis for DENV infection includes detection of the virus, genome, non-structural (NS)-1 antigen or IgM/IgG antibodies, or a combination of these tests [4]. NS1 is a highly conserved glycoprotein of flaviviruses that can be detected in blood samples, most often between one and nine days after the onset of symptoms, which is very efficient for early diagnosis of DENV infection [5]. According to the WHO recommendations, confirmatory diagnosis of DENV infection includes virus detection by PCR or virus culture, detection of IgM seroconversion in paired sera, IgG seroconversion, or ≥four-fold increase in the IgG titer in paired sera [1]. ELISA-based serological tests can detect IgM, IgG, or the NS1 glycoprotein [6]. As many patients seek medical care five days after fever onset, anti-DENV IgM/IgG become suitable markers for diagnosing a recent DENV infection, and the anti-DENV IgG test can help differentiate primary and secondary DENV infections [7]. In addition, rapid diagnostic tests (RDTs) are commonly used for DENV detection because of their simplicity and rapidity [3]. Several ELISAs and RDTs are now widely available from different manufacturers. However, independent validation and comparative evaluation remain limited. Thus, we compared the performance of six commercial serological tests including three RDTs for diagnosing DENV infection. This study is the first to carry out such a comparison.
We tested a total of 138 single blood samples, including 34 samples from Korean patients suspected for DENV infection, 60 from patients with confirmed DENV infection (purchased from TRINA BIOREACTIVES AG, Nänikon, Switzerland), and 44 from healthy Korean subjects in a dengue non-endemic area. The supplier reported that the 60 samples were confirmed by clinical diagnosis and the DENV IgM test. Serum samples from Korean patients were sent to the laboratory of Seoul St. Mary's Hospital, Seoul, Korea, and stored at −80℃ until testing. The Institutional Review Board of Seoul St. Mary's Hospital approved this study (XC16SNMI0049K, KC17SNSI0246). Informed consent was waived because the current study was performed using leftover blood samples.
Evaluation was performed with three sets of ELISAs and three RDTs: (i) DENV Detect NS1 ELISA, DENV Detect IgM capture ELISA and DENV Detect IgG ELISA (InBios International, Inc., Seattle, WA, USA); (ii) Anti-Dengue virus IgM Human ELISA and Anti-Dengue virus IgG Human ELISA (Abcam, Cambridge, MA, USA); (iii) Dengue virus NS1 ELISA, Anti-Dengue virus ELISA (IgM) and Anti-Dengue virus ELISA (IgG) (Euroimmun, Lübeck, Germany); (iv) Asan Easy Test Dengue NS1 Ag 100 and Asan Easy Test Dengue IgG/IgM (Asan Pharm, Seoul, Korea); (v) SD BIOLINE Dengue Duo (Standard Diagnostics Inc., Seoul, Korea); and (vi) Ichroma Dengue NS1 and Ichroma Dengue IgG/IgM (Boditech Med, Chuncheon, Korea). All six tests can detect all three targets (NS1 antigen, DENV IgM antibody, and DENV IgG antibody) except for Abcam ELISA, which can detect only the anti-DENV IgM and IgG antibodies. All tests were performed according to the manufacturers' instructions. We performed the tests only once and repeated the tests when there was an invalid result, according to the manufacturers' instructions. The test characteristics are summarized in Table 1. For the NS1 antigen and DENV IgM/IgG tests, cross-reactivity was tested using samples from patients with confirmed chikungunya infection (N=20) (TRINA BIOREACTIVES AG). To detect cross-reactivity with other viral infections, serum from patients with positive IgG against Epstein–Barr virus (N=10), cytomegalovirus (N=10), hepatitis A virus (N=10), hepatitis B virus (N=10), family
For each NS1 antigen, DENV IgM, and DENV IgG test, reference positive results were defined when the samples were positive using two or more of the ELISAs examined. All samples that were not classified as reference positives were considered as reference DENV negative samples. The reference results for NS1 antigen, IgM, IgG, and PCR of all clinical samples are provided in Supplemental Data Table S1.
Statistical analyses were performed using MedCalc version 15.5 (MedCalc, Ostend, Belgium). Diagnostic accuracy, agreement rates, and Cohen's kappa coefficients (κ) between tests were calculated, as well as sensitivity, specificity, positive predictive value (PPV) and negative predictive value (NPV) for each test. The κ values were shown with 95% confidence interval and were interpreted as very good (0.81–1.00), good (0.61–0.80), moderate (0.41–0.60), fair (0.21–0.40), or poor (<0.20). Pearson's correlation coefficient was calculated between optical density (OD) values of Euroimmun with the other tests.
For NS1 antigen detection, the InBios and Euroimmun ELISAs showed 100% sensitivity and 99.2% specificity. The three RDTs (Asan, SD, and Boditech Med) showed high specificity (99.2–100.0%), PPV (85.7–100%) and NPV (93.9–96.1%). However, the sensitivities of these RDTs were low (42.9–64.3%; Table 2). Previous studies have reported a wide range of performances (sensitivity: 18.6–96.9%; specificity: 53–100%) [8,9].
For DENV IgM detection, the six tests showed variable sensitivities (38.1–90.5%). Two ELISAs (InBios and Abcam) and one RDT (Boditech Med) had higher sensitivities (81.0–90.5%) than the other three tests (Euroimmun, Asan, and SD; 38.1–49.2%). When we evaluated the combined NS1 antigen and DENV IgM antibody results for each test, five of the six tests (except Abcam) showed variable sensitivity (44.1–95.6%) and NPV (64.5–95.8%). However, they showed high specificity (92.9–100%) and high PPV (92.3–100%). Of these, InBios ELISA showed a sensitivity of 95.6% and a specificity of 97.1% for NS1+IgM results, while Euroimmun, Asan, and SD ELISAs showed lower sensitivity. When the NS1+IgM+IgG results were combined, the diagnostic accuracy of all the RDTs improved.
The sensitivity of DENV IgM antibodies is lower for secondary infections than for primary infections, and the DENV IgM antibody alone does not provide sufficient results for DENV diagnosis [7,10]. Therefore, a combination of NS1 and IgM antibody tests is recommended [8]. When we combined the NS1 antigen and IgM antibody results, the RDTs and ELISAs had similar performance, confirming previous reports that RDTs could be useful in clinical settings [3,11,12,13]. In our study, the sensitivity of the Euroimmun ELISA for IgM detection was as low as 38.1%. However, the correlations between OD values measured by Euroimmun and those by the other two ELISAs were strong (Abcam: r=0.94,
For IgG antibody detection, all three ELISAs and one RDT (Boditech Med) showed sensitivities >94.0%. Previous studies have reported variable sensitivities for DENV IgG and IgM ELISAs (7.8–88.9% and 20.2–99%, respectively) with various specificities (63.5–100% and 52–100%, respectively) [4,9]. However, the RDTs showed relatively low NPVs (73.9–94.3%) than the ELISAs (94.4–100.0%).
In the evaluation of agreement, the InBios and Euroimmun ELISAs showed a 98.6% agreement rate for NS1 antigen detection (κ=0.93) (Table 3). The three RDTs also showed a high agreement rate of 97.1–97.8% for NS1 detection (κ=0.73–0.82). Although we evaluated NS1 antigen tests based on reference serology instead of clinical diagnosis, the agreements between tests were good, suggesting that NS1 antigen tests could support diagnostic tests for acute DENV infection. For DENV IgM detection, the InBios and Abcam ELISA results showed good agreement (86.2%, κ=0.71), while Euroimmun ELISA showed fair (70.3%, κ=0.34) and moderate (79.7%, κ=0.52) agreement with the InBios and Abcam IgM tests, respectively. When agreements between RDT and ELISA DENV IgM tests were compared, the Boditech Med IgM test results showed good agreement with InBios (84.8%, κ=0.69) or Abcam (81.2%, κ=0.61). These six tests for DENV IgG antibody demonstrated very good to good agreement (80.4–97.1%, κ>0.60) except between Abcam ELISA IgG and the Asan RDT IgG test (77.5%, κ=0.57).
When cross-reactivity was tested using samples from patients with confirmed infection with chikungunya virus and hepatitis C virus, none of the five DENV NS1 antigen tests showed positivity. However, all six DENV IgG tests demonstrated variable potential cross-reactivity (range, 30–90%) with chikungunya IgG (Supplemental Data Fig. S1). High intra-genus cross-reactivity with
This study had some limitations. We could not evaluate the results by DENV serotype or primary/secondary infection status and could not test cross-reactivation with other common
Over the years, immunoassays have progressed dramatically from manual to automated bench systems. However, simple manual immunoassays or point-of-care tests are still desirable in regions with insufficient laboratory resources and expertise [16]. The global spread of DENV and the growing number of international travelers have necessitated that attention be paid to DENV infection in non-endemic areas, including Korea [17].
In conclusion, the ELISAs demonstrated good performance for diagnosing DENV infection, and the RDTs showed reliable results compared with ELISA. However, clinical laboratories should be aware of performance variations across tests and the possibilities of cross-reactivity.
NS1 antigen, IgM, IgG, and PCR reference results of all clinical samples
alm-39-566-s001.pdfFalse positive rate of six DENV IgG tests for samples with IgG against other viruses; DENV Detect IgG ELISA (InBios International, Inc., Seattle, WA, USA); Anti-Dengue virus IgG Human ELISA (Abcam, Cambridge, MA, USA); Anti-Dengue virus ELISA (IgG) (Euroimmun, Lübeck, Germany); Asan Easy Test Dengue IgG/IgM (Asan Pharm, Seoul, Korea); SD BIOLINE Dengue Duo (Standard Diagnostics Inc., Seoul, Korea); Ichroma Dengue IgG/IgM (Boditech Med, Chuncheon, Korea).
Abbreviations: EBV, Epstein–Barr virus; DENV, dengue virus; CMV, cytomegalovirus; HAV, hepatitis A virus; HBV, hepatitis B virus; HCV, hepatitis C virus.
alm-39-566-s002.pdf