Detection of a High-Dose Hook Effect and Evaluation of Dilutions of Urine Myoglobin Specimens Using a Serum Myoglobin Assay
Joshua Joon Hyung Hunsaker , B.S.1, Sonia Leilani La’ulu , M.B.A.1, Emily Zupan, B.S.2, Dhwani Patel , B.S.2, Vrajesh Pandya , Ph.D.2,3, and Joseph William Rudolf, M.D.2,3
1ARUP Institute for Clinical and Experimental Pathology, Salt Lake City, UT, USA; 2ARUP Laboratories, Salt Lake City, UT, USA; 3Department of Pathology, University of Utah Health, Salt Lake City, UT, USA
Correspondence to: Joseph William Rudolf, M.D.
Department of Pathology, ARUP Laboratories, University of Utah, 500 Chipeta Way, Salt Lake City, UT 84108, USA
E-mail: joseph.rudolf@path.utah.edu
This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
Dear Editor,
Severe muscle injuries can lead to the release of myoglobin into the bloodstream, subsequently filtering into the urine via the kidneys [1, 2]. Elevated myoglobin levels can be detrimental to the kidneys, causing renal damage or failure. Therefore, urinary myoglobin measurement is crucial; however, existing commercially available kits are designed for serum or plasma [3, 4]. High myoglobin concentrations in serum or plasma can induce a high-dose hook effect [5, 6]. Using a serum or plasma myoglobin assay for urine specimens poses a risk of a hook effect owing to high myoglobin concentrations that may be present in urine from patients with severe muscle injuries [1, 7]. Therefore, the predilution of urine specimens would address high concentrations and potential hook effect but may impact accuracy as diluted specimens might approach the assay sensitivity [4].
Various urinary myoglobin dilutions were evaluated using the Elecsys Myoglobin assay (Cat#. 12178214160; Roche Diagnostics; Indianapolis, IN, USA) on a cobas e602 analyzer, which has an analytical measuring interval (AMI) of 1.18–168.54 nmol/L (21–3,000 ng/mL). Results were reported in ng/mL and converted to the International System of Units based on the molecular weight of myoglobin (17.8 kD) [8]. Residual, de-identified, frozen (−20°C) urine specimens were used (IRB protocol #00007275; informed consent waived). Specimens were pH-adjusted to 8–9 at the time of collection using Na2CO3. Specimens were diluted onboard with Elecsys Diluent Universal (Roche Diagnostics), and dilutions of 1:400 were previously investigated in accordance with CLSI guidelines (CLSI EP34) [9].
As a preliminary assessment, urine specimens (N=7; 56.19–1,516.85 nmol/L) were retested without dilution (neat) and at a 1:50 dilution. Of these, one produced a suspected hook effect, where the neat result was 100.00 nmol/L but was 1,496.46 nmol/L at a 1:50 dilution. This suggests that a hook effect in urine may occur at lower analyte concentrations than specified for the serum matrix by the manufacturer, as a hook effect does not occur for serum myoglobin concentrations up to 1,685.39 nmol/L [4]. For all remaining specimens (N=6; 23.76–68.03 nmol/L), the results were elevated by approximately 2.5-fold when using a 1:50 dilution, suggesting that the specimens had concentrations low enough to impede a dilution of that magnitude, potentially affecting measurement accuracy.
To investigate whether a lower dilution factor would allow the detection of a hook effect while providing more accurate results, additional urine specimens (N=35; 56.19–33,539.33 nmol/L) were reanalyzed neat, 1:10, and 1:50 dilutions. Specimens below the AMI (<168.54 nmol/L) were categorized as not exhibiting a hook effect. Specimens that produced results above the AMI after a 1:50 dilution (>8,426.97 nmol/L) were further diluted at 1:400. A hook effect was detected in 30 of the 35 specimens at both the 1:10 and 1:50 dilutions, with several specimens requiring a 1:400 dilution to obtain a result (Table 1).
-
Table 1 . Detection of a high-dose hook effect using 1:10, 1:50, and 1:400 dilutions
Sample number Neat
(nmol/L)1:10 dilution
(nmol/L)Hook effect detected
(1:10; >168.54 nmol/L)1:50 dilution
(nmol/L)Hook effect detected
(1:50; >168.54 nmol/L)1:400 dilution
(nmol/L)1 48.89 125.79 No hook effect 154.83 No hook effect 2 41.63 1,049.21 Detected 2,412.81 Detected 3 10.15 274.04 Detected 2,901.46 Detected 4 85.34 74.78 No hook effect 85.45 No hook effect 5 55.85 197.98 Detected 247.36 Detected 6 64.27 >1,685.40 Detected 2,429.38 Detected 7 42.74 >1,685.40 Detected >8,426.97 Detected 21,166.57 8 25.99 80.62 No hook effect 115.79 No hook effect 9 110.28 1,140.84 Detected 1,316.69 Detected 10 43.51 >1,685.40 Detected 6,093.54 Detected 11 20.08 732.08 Detected 5,770.34 Detected 12 89.61 111.97 No hook effect 110.79 No hook effect 13 20.31 282.58 Detected 530.56 Detected 14 30.52 >1,685.40 Detected >8,426.97 Detected 31,265.56 15 34.17 459.78 Detected 824.10 Detected 16 28.51 >1,685.40 Detected >8,426.97 Detected 12,656.46 17 34.17 >1,685.40 Detected >8,426.97 Detected 15,379.49 18 9.56 901.91 Detected >8,426.97 Detected 21,466.80 19 28.60 1,326.57 Detected 6,596.07 Detected 20 >168.54 >1,685.40 Detected 3,221.52 Detected 21 31.82 1,133.48 Detected 5,771.01 Detected 22 48.18 952.13 Detected 2,553.48 Detected 23 16.34 384.89 Detected 3,048.03 Detected 24 61.57 1,084.78 Detected 1,499.72 Detected 25 8.31 302.87 Detected 2,155.17 Detected 26 28.19 413.99 Detected 613.09 Detected 27 >168.54 >1,685.40 Detected 2,070.11 Detected 28 7.66 265.34 Detected 390.22 Detected 29 10.99 75.96 No hook effect 152.87 No hook effect 30 63.43 >1,685.40 Detected 6,177.75 Detected 31 96.24 395.45 Detected 431.01 Detected 32 31.87 552.81 Detected 1,724.72 Detected 33 77.53 1,501.52 Detected 2,090.51 Detected 34 108.71 224.44 Detected 259.21 Detected 35 >168.54 1,316.69 Detected 1,617.02 Detected
Further, we performed a recovery experiment by serially diluting a high-concentration urine specimen (48,744.33 nmol/L) with a low-concentration urine specimen (6.18 nmol/L). Expected concentrations were calculated based on the high- and low-concentration specimens used. Each serially diluted specimen was tested neat and at 1:10, 1:50, and 1:400 dilutions, and the results were compared to the expected concentrations. However, the dilutions were only performed when the anticipated concentration was within the AMI of the assay (1.18–168.54 nmol/L). The acceptable limit of recovery (80.4%–119.6%) was based on a previously reported total allowable error for myoglobin in serum [10]. For each of the 1:10, 1:50, and 1:400 dilutions, recoveries exceeded 119.6% when the diluted specimen result (prior to correcting for the dilution factor) was 4.06–37.29 nmol/L (Table 2). The manufacturer recommends that the serum concentration after dilution should exceed 2.81 nmol/L; nevertheless, we observed that a higher concentration may be necessary for diluted urine specimens. Diluted results were within acceptable limits of recovery when the diluted specimen produced results between 35.86–132.19 nmol/L (Table 2). These findings demonstrate that accuracy reduces with decreasing myoglobin concentrations owing to the dilution process. Therefore, a smaller predilution factor (1:10) is recommended with additional dilutions (1:50 and 1:400) as necessary to detect a hook effect while avoiding inaccuracies for low-concentration specimens.
-
Table 2 . Evaluation of dilutions using the Roche Myoglobin assay on the cobas e602 analyzer
Expected concentration (nmol/L) Result of diluted specimen* (nmol/L) Dilution Final reported result* (nmol/L) % Recovery 48,744.33 121.86 1:400 48,744.33 100.0 24,375.28 68.00 1:400 27,201.07 111.6 12,190.73 35.86 1:400 14,342.58 117.7 6,098.48 132.19 1:50 6,609.66 108.4 3,052.36 67.31 1:50 3,365.67 110.3 1,529.27 37.29 1:50 1,864.61 121.9 767.75 84.94 1:10 849.44 110.6 386.97 44.03 1:10 440.34 113.8 196.57 23.51 1:10 235.11 119.6 101.40 12.90 1:10 129.04 127.3 53.82 6.75 1:10 67.53 125.5 30.00 4.06 1:10 40.56 135.2 18.09 Not applicable No dilution 18.93 104.6 12.13 Not applicable No dilution 12.87 106.1 9.16 Not applicable No dilution 9.61 104.9 7.70 Not applicable No dilution 7.87 102.2 6.97 Not applicable No dilution 7.08 101.6 *The result of the diluted specimen is calculated by dividing the final reported result divided by the dilution performed (e.g., if the final reported result were 500 nmol/L and the dilution performed was 1:10, the calculation would be 500 nmol/L/10=50 nmol/L).
Overall, these findings support a predilution step when analyzing urine specimens using the Roche Myoglobin assay. A hook effect was observed in urine at lower concentrations than the serum hook effect threshold reported in the assay package insert. We conclude that to effectively use a serum myoglobin assay to measure myoglobin in urine, an appropriate predilution factor should be carefully validated to enable the detection of a high-dose hook effect while producing results with good accuracy and clinical acceptability.
ACKNOWLEDGEMENTS
The authors gratefully acknowledge Taylor M. Snow for specimen acquisition.
AUTHOR CONTRIBUTIONS
Hunsaker JJH performed the experimental work, analyzed the data, and drafted the manuscript. All authors contributed to the study conception and design, were involved in clinical evaluation and result interpretation, and read and approved the final manuscript.
CONFLICTS OF INTEREST
None declared.
RESEARCH FUNDING
This study was supported by the ARUP Institute for Clinical and Experimental Pathology.
References
- Anwar MY, Gupta V. Myoglobinuria. StatPearls. Treasure Island (FL): StatPearls Publishing, 2022: 1-3.
- Chen-Levy Z, Wener MH, Toivola B, Daum P, Reyes M, Fine JS. Factors affecting urinary myoglobin stability in vitro. Am J Clin Pathol 2005;123:432-8.
- de Waard H, van 't Sant P. Use of serum myoglobin assays for urine myoglobin measurements can cause false-negative results. Clin Chem 2009;55:1240-1.
- Elecsys Myoglobin Package Insert. 2022-10, v.8.0. Roche Diagnostics, Indianapolis, IN, USA.
- Kurt-Mangold M, Drees D, Krasowski MD. Extremely high myoglobin plasma concentrations producing hook effect in a critically ill patient. Clin Chim Acta 2012;414:179-81.
- Kikuchi H, Ohta A, Takahashi Y, Shibata A, Ohtake T, Watanabe K. High-dose hook effect in an immunochromatography-optical quantitative reader method for myoglobin. Clin Chem 2003;49:1709-10.
- Tate J, Ward G. Interferences in immunoassay. Clin Biochem Rev 2004;25:105-20.
- Lamb EJ, Jones GRD. Kidney function tests. In: Rifai N, Chiu RWK, et al, eds. Tietz textbook of laboratory medicine. 7th ed. Saint-Louis, MO: Elsevier, 2022:1531-91.
- CLSI. Validating an extended measuring interval. Establishing and verifying an extended measuring interval through specimen dilution and spiking. 1st ed. CLSI EP34. Wayne, PA: Clinical and Laboratory Standards Institute, 2018:46-55.
- Ricós C, Alvarez V, Cava F, García-Lario JV, Hernández A, Jiménez CV, et al. Current databases on biological variation: pros, cons and progress. Scand J Clin Lab Invest 1999;59:491-500.