Interinstitutional Comparison of Vancomycin Area Under the Concentration–Time Curve Estimation in Korea: Need for Standardized Operational Protocols for Therapeutic Drug Monitoring Consultation
2025; 45(1): 85-89
Ann Lab Med 2021; 41(3): 268-276
Published online May 1, 2021 https://doi.org/10.3343/alm.2021.41.3.268
Copyright © Korean Society for Laboratory Medicine.
Hyun-Ki Kim , M.D.1, Hyung-Doo Park , M.D.2, Sang-Guk Lee , M.D.3, Hyojin Chae , M.D.4, Sang Hoon Song , M.D.5, Yong-Wha Lee , M.D.6, Yeo-Min Yun , M.D.7, Sunhyun Ahn , M.D.8, Serim Kim , M.D.9, Sun Min Lee , M.D.10, Soo-Youn Lee , M.D.2, and Sail Chun, M.D.11 ; on behalf of the Clinical Mass Spectrometry Research Committee of the Korean Society of Clinical Chemistry
1Department of Laboratory Medicine, University of Ulsan College of Medicine, Ulsan University Hospital, Ulsan, Korea; 2Department of Laboratory Medicine and Genetics, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Korea; 3Department of Laboratory Medicine, Severance Hospital, Yonsei University College of Medicine, Seoul, Korea; 4Department of Laboratory Medicine, College of Medicine, The Catholic University of Korea, Seoul, Korea; 5Department of Laboratory Medicine, Seoul National University Hospital and College of Medicine, Seoul, Korea; 6Department of Laboratory Medicine and Genetics, Soonchunhyang University Bucheon Hospital, Soonchunhyang University College of Medicine, Bucheon, Korea; 7Department of Laboratory Medicine, Konkuk University School of Medicine, Seoul, Korea; 8Seoul Clinical Laboratories, Yongin, Korea; 9Green Cross Laboratories, Yongin, Korea; 10Department of Laboratory Medicine, Pusan National University Yangsan Hospital, Yangsan, Korea; 11Department of Laboratory Medicine, University of Ulsan College of Medicine and Asan Medical Center, Seoul, Korea
Correspondence to: Sail Chun, M.D.
Department of Laboratory Medicine, University of Ulsan College of Medicine and Asan Medical Center, 88 Olympic-ro 43-gil, Songpa-gu, Seoul 05505, Korea
Tel: +82-2-3010-4513
Fax: +82-2-478-0884
E-mail: sailchun@amc.seoul.kr
Background: Liquid chromatography coupled to tandem mass spectrometry (LC-MS/MS) is increasingly used for immunosuppressive drug tests. However, most LC-MS/MS tests are laboratory-developed and their agreement is unknown in different Korean laboratories. This interlaboratory comparison study evaluated test reproducibility and identified potential error sources.
Methods: Test samples containing three concentrations of tacrolimus, sirolimus, everolimus, cyclosporine, and mycophenolic acid were prepared by pooling surplus samples from patients undergoing routine therapeutic drug monitoring and tested in duplicate in the participating 10 clinical laboratories. Reconstitution and storage experiments were conducted for the commonly used commercial calibrator set. The robust estimators of reproducibility parameters were calculated. Spearman’s rank correlation coefficient (rho, ρ) was used to evaluate the correlation between drugs. Multiple linear regression was used to determine whether the experimental conditions alter the calibration curves.
Results: The reproducibility coefficient of variation exceeded 10% only for sirolimus concentrations 1 and 2 (10.8% and 12.5%, respectively) and everolimus concentrations 1 and 2 (12.3% and 11.4%, respectively). The percent difference values showed weak correlations between sirolimus and everolimus (ρ=0.334, P =0.175). The everolimus calibration curve slope was significantly altered after reconstitution following prolonged 5°C storage (P =0.015 for 14 days; P =0.025 for 28 days); the expected differences at 6 ng/mL were 0.598% for 14 days and 0.384% for 28 days.
Conclusions: LC-MS/MS test reproducibility for immunosuppressive drugs seems to be good in the Korean clinical laboratories. Continuous efforts are required to achieve test standardization and harmonization, especially for sirolimus and everolimus.
Keywords: Immunosuppressive drug, Therapeutic drug monitoring, Liquid chromatography coupled to tandem mass spectrometry, Interlaboratory comparison, Sirolimus, Everolimus, Reproducibility
Immunosuppressive drugs are essential for successful transplantation, as they reduce the risk of rejection [1]. The five immunosuppressive drugs most commonly prescribed for the prevention of graft loss and organ rejection are tacrolimus (TAC), sirolimus (SIR), everolimus (EVE), cyclosporine (CSA), and mycophenolic acid (MPA) [2, 3]. Calcineurin inhibitors, such as TAC, SIR, EVE, and CSA, have narrow therapeutic ranges and high interindividual pharmacokinetic variability. Therefore, therapeutic drug monitoring (TDM) is widely practiced to determine the dosage adjustments needed to reach target blood concentrations [4]. Although the utility of TDM for MPA remains controversial, monitoring MPA concentrations is also sometimes recommended [5, 6].
Reliable measurements are essential for effective TDM of immunosuppressive drugs. Liquid chromatography coupled to tandem mass spectrometry (LC-MS/MS) is increasingly used for TDM of immunosuppressive drugs, mainly owing to its high analytical specificity and the ability to perform multiplex tests [1]. However, most LC-MS/MS tests used in laboratories are developed, calibrated, and validated in-house. This could lead to relatively poor interlaboratory agreement between independently developed tests, thus hindering the adoption of common clinical practice guidelines [7].
There are proficiency testing (PT) programs for immunosuppressive drug TDM operated by the College of American Pathologists (CAP) or the Korean Association of External Quality Assessment Service (KEQAS). However, the CAP international PT programs do not provide nation-specific information [8], and the KEQAS PT programs included only a few participating laboratories using LC-MS/MS methods [9]. Therefore, the test reproducibility of immunosuppressive drug TDM by LC-MS/MS in Korea is unknown. An interlaboratory comparison study could determine test reproducibility across laboratories, as well as overall repeatability [10]. Interlaboratory comparisons could also contribute to laboratory performance improvement, with proper reviews of test practices and follow-up on the implementation of corrective actions [11-13].
We conducted an interlaboratory comparison study by including most of the clinical laboratories performing immunosuppressive drug TDM by LC-MS/MS in Korea and investigated calibrator reconstitution and storage conditions as potential error sources. We aimed to document current practices used for immunosuppressive drug TDM by LC-MS/MS, evaluate the reproducibility of these tests, and identify potential error sources contributing to interlaboratory imprecision.
Based on the results of a previous survey of members of the Korean Society of Clinical Chemistry [14], 10 clinical laboratories performing immunosuppressive drug TDM by LC-MS/MS were asked to participate in this study. After enrollment, the test samples were prepared in Asan Medical Center, Seoul, Korea, dispatched to the participating laboratories on dry ice in a Styrofoam box on 23 July 2018, and reached the laboratories within two days. Participating laboratories were asked to perform the test within three days of receiving the test samples. For each test sample, participating laboratories were asked to make duplicate aliquots for testing before the sample preparation step. Test results and method information, including the dates of reception and test, calibrator, internal standard, LC manufacturer, MS manufacturer and model, and extraction method, were collected in a formatted Excel file. Additional surveys regarding calibrator reconstitution protocols were conducted by e-mail. The data was compiled and analyzed at Asan Medical Center. The Institutional Review Board of Asan Medical Center exempted the approval for this study (reference number: 2018-1070).
Surplus samples from patients undergoing routine TDM were pooled to achieve three different concentrations of test samples for each drug. The samples included EDTA whole blood for TAC, SIR, EVE, and CSA and EDTA plasma for MPA. The samples were collected for seven days and stored at 4°C up to seven days before pooling. The samples were pooled based on their reported concentrations to achieve concentrations below, within, and above the therapeutic range of each drug. Each 1 mL pooled sample was distributed in a glass vial after mixing on a roller mixer for 1 hour. The minimum number of vials for each test sample was 10; for TAC concentration 2 (TAC-2), 30 vials were prepared to evaluate homogeneity and short-term stability. The test samples were stored at -70°C before delivery. Thus, the samples underwent only one freeze-thaw cycle before being tested at the participating laboratories. For TAC-2, 10 vials were tested in triplicate on the day of production to evaluate homogeneity. To evaluate short-term stability, two vials were stored at 5°C for five days and then tested for TAC in duplicate. These four results were compared with the homogeneity values. In the homogeneity assessment, a between-sample standard deviation < 0.3
All test samples were in a frozen state on arrival at each laboratory. Details of the methods used in each laboratory are shown in Table 1. All laboratories conducted simultaneous detection with the following combinations: TAC+SIR+EVE+CSA (six laboratories) and TAC+SIR+EVE (three laboratories); one laboratory (Lab F) that was supposed to perform a CSA test with a TAC+SIR+EVE+CSA combination did not receive the CSA test samples due to a communication error. In total, TAC detection was performed in 10 laboratories, SIR detection in nine laboratories, EVE detection in nine laboratories, CSA detection in eight laboratories, and MPA detection in three laboratories. All participating laboratories used commercial calibrator sets from Chromsystems Instruments & Chemicals GmbH (Gräfelfing, Germany).
Drug measurement methods used in participating laboratories
Laboratory code | Test drugs | Internal standard | LC manufacturer* | MS instrument manufacturer* | MS instrument model |
---|---|---|---|---|---|
A | TAC+SIR+EVE | Ascomycin+SIR-d3+EVE-d4 | Shimadzu | SCIEX | Triple Quad 4500 |
CSA | CSD | Agilent | Agilent | 6460 Triple Quadrupole | |
MPA | Indomethacin | Waters | Waters | Quattro Micro | |
B | TAC+SIR+EVE+CSA | Ascomycin | Waters | Waters | Quattro Premier |
C | TAC+SIR+EVE | Ascomycin+SIR-d3+EVE-d4 | Waters | Waters | Xevo TQ-S |
D | TAC+CSA | Ascomycin+CSD | Waters | Waters | Xevo TQD |
MPA | Indomethacin | Waters | Waters | Xevo TQD | |
E | TAC+SIR+EVE | Ascomycin | Waters | Waters | Xevo TQD |
CSA | CSD | Waters | Waters | Xevo TQD | |
F | TAC+SIR+EVE+CSA | Ascomycin+EVE-d4+CSD | Agilent | Agilent | 6490 Triple Quadrupole |
G | TAC+SIR+EVE+CSA | Ascomycin+SIR-d3+EVE-d4+CSD | Agilent | SCIEX | API 4000 |
H | TAC+SIR+EVE+CSA | TAC-13C,d2+SIR-d3+EVE-d4+CSA-d4 | Agilent | SCIEX | API 3200 |
I | TAC+SIR+EVE+CSA | Ascomycin | Agilent | SCIEX | QTRAP 5500 |
MPA | MPA-d3 | Agilent | SCIEX | QTRAP 5500 | |
J | TAC+SIR+EVE+CSA | Ascomycin+CSD | Agilent | SCIEX | Triple Quad 3500 |
Most laboratories used precipitation with organic solvent mixture followed by centrifugation for extraction, except Lab I, which used liquid/liquid extraction. All the participating laboratories used the commercial calibrator from Chromsystems Instruments & Chemicals GmbH (Gräfelfing, Germany).
*The manufacturer names are indicated using their short names: Agilent, Agilent Technologies (Santa Clara, CA, USA); SCIEX, AB SCIEX (Foster City, CA,USA); Shimadzu, Shimadzu Corporation (Kyoto, Japan); Waters, Waters Corporation (Milford, MA, USA).
Abbreviations: CSA, cyclosporine; CSD, cyclosporine D; EVE, everolimus; LC, liquid chromatography; MPA, mycophenolic acid; MS, mass spectrometry;TAC, tacrolimus; SIR, sirolimus.
Based on the survey results of the calibrator reconstitution methods of participating laboratories, we conducted an experiment on calibrator reconstitution and storage conditions. To evaluate the impact of the calibrator reconstitution process and prolonged storage after reconstitution, we used the lyophilized calibrator 6PLUS1 Multilevel Whole Blood Calibrator Set (Lot no. 4917; Chromsystems Instruments & Chemicals GmbH) with six different concentrations plus one blank value. The calibrator contained TAC, SIR, EVE, and CSA; however, CSA could not be measured in the experimental test system.
The calibrators were stored at -70°C until use and then thawed at room temperature (20–25°C) for 30 minutes before adding 2.0 mL of distilled water to each vial. The contents of the vials were then mixed under two different conditions: static incubation (sitting) for 30 minutes at room temperature (20–25°C) after the addition of water followed by mixing on a roller for 60 minutes (Condition A0) or mixing on a roller for 20 minutes without static incubation (Condition B0). Aliquots from these vials were tested in four replicates. Aliquots from Condition A0 vials were initially frozen at -20°C and then thawed on different days, resulting in four storage conditions: 42 days at -20°C (Condition A1), 35 days at -20°C and seven days at 5°C (Condition A2), 28 days at -20°C and 14 days at 5°C (Condition A3), and 14 days at -20°C and 28 days at 5°C (Condition A4). These aliquots were tested in duplicate on the same day.
Simultaneous quantification of TAC, SIR, and EVE was performed for the calibrators in the same manner as for patient samples at Asan Medical Center. Briefly, 40 µL samples were prepared by protein precipitation using 80 µL of aqueous 0.1 M zinc sulfate and 200 µL of acetonitrile containing the internal standards. The internal standard solution consisted of ascomycin (Cerilliant, Round Rock, TX, USA), SIR-d3 (IsoSciences, Ambler, PA, USA), and EVE-d4 (Cerilliant). HPLC was conducted using an ACQUITY UPLC I-Class system (Waters Corporation, Milford, MA, USA) equipped with an ACQUITY UPLC HSS SB C18 column (30 mm × 2.1 mm, 4 µm; Waters Corporation). The system was operated at a flow rate of 0.2 mL/min with a column temperature of 55°C. The mobile phase for isocratic elution consisted of 50% methanol (HPLC grade; Merck, Darmstadt, Germany) and 50% ammonium acetate buffer (2 mM, pH 2.7, with ≥ 98% acetate ammonium; Sigma-Aldrich, St. Louis, MO, USA). MS was detected on a Xevo TQ-S LC-MS/MS system (Waters Corporation) using positive electrospray ionization and argon as the collision gas. Peak areas were obtained by multiple reaction monitoring of the following mass transitions for quantification: TAC (m/z 821.5 > 768.5), SIR (m/z 931.6 > 864.6), EVE (m/z 975.7 > 908.6), ascomycin (m/z 809.5 > 756.5), SIR-d3 (m/z 934.6 > 864.6), and EVE-d4 (m/z 979.7 > 912.6). Response values were calculated using the resultant peak areas (area of analyte/area of internal standard). LC-MS/MS analysis and data acquisition were managed using Masslynx software version 4.1 (Waters Corporation). TargetLynx software version 4.1 (Waters Corporation) was used for chromatographic peak detection and baseline determination.
Data from the participating laboratories were used to estimate test reproducibility. The repeatability standard deviation (Sr) is the root mean square value of all differences in
The standard deviation of the sample averages The standard deviation of the sample averages (
where
When
The reproducibility coefficient of variation (
Robust statistical methods based on the medians were also used to describe the result distribution. The robust repeatability standard deviation (
which is the scaled median absolute deviation (
For the result analysis, a multiple linear regression model was used to compare the slopes of the calibration curves (non-weighted):
where the response value is
In the homogeneity test, the overall average of TAC-2 was 8.99 ng/mL. Because the variance of sample averages was < 1/3 of the within-sample variance, the between-sample standard deviation was estimated as 0. Therefore, the homogeneity was considered sufficient. In the short-term stability test, the overall average of TAC-2 was 8.88 ng/mL. Its difference from the homogeneity mean was 0.11 ng/mL, which was lower than the calculated check value (0.17, 0.3
Statistical summary of the laboratory measurement results is shown in Table 2. The robust reproducibility CVS (
Summary statistics of laboratory measurement results
Drug | Participating laboratories (N) | Test sample | General without outlier removal | Robust statistics | ||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|
Mean ( | Median ( | |||||||||||
TAC (ng/mL) | 10 | TAC-1 | 5.41 | 0.41 | 0.48 | 0.57 | 10.5 | 5.59 | 0.31 | 0.24 | 0.32 | 5.8 |
TAC-2 | 8.93 | 0.39 | 0.57 | 0.63 | 7.1 | 8.84 | 0.24 | 0.19 | 0.25 | 2.8 | ||
TAC-3 | 15.83 | 0.50 | 1.33 | 1.37 | 8.7 | 16.05 | 0.30 | 1.43 | 1.45 | 9.0 | ||
SIR (ng/mL) | 9 | SIR-1 | 3.50 | 0.19 | 0.29 | 0.32 | 9.1 | 3.55 | 0.14 | 0.37 | 0.38 | 10.8 |
SIR-2 | 5.83 | 0.36 | 0.88 | 0.92 | 15.7 | 5.55 | 0.27 | 0.67 | 0.69 | 12.5 | ||
SIR-3 | 12.08 | 1.13 | 0.61 | 1.00 | 8.3 | 11.96 | 0.85 | 0.61 | 0.86 | 7.2 | ||
EVE (ng/mL) | 9 | EVE-1 | 2.43 | 0.25 | 0.33 | 0.38 | 15.5 | 2.45 | 0.19 | 0.27 | 0.30 | 12.3 |
EVE-2 | 3.53 | 0.39 | 0.36 | 0.46 | 12.9 | 3.55 | 0.27 | 0.36 | 0.40 | 11.4 | ||
EVE-3 | 5.89 | 0.25 | 0.49 | 0.53 | 8.9 | 5.90 | 0.19 | 0.52 | 0.54 | 9.1 | ||
CSA (ng/mL) | 8 | CSA-1 | 30.10 | 1.67 | 2.83 | 3.06 | 10.2 | 29.85 | 0.78 | 0.70 | 0.89 | 3.0 |
CSA-2 | 63.78 | 2.32 | 3.77 | 4.12 | 6.5 | 63.45 | 1.15 | 2.37 | 2.51 | 4.0 | ||
CSA-3 | 149.93 | 4.55 | 10.44 | 10.92 | 7.3 | 149.40 | 3.53 | 9.01 | 9.35 | 6.3 | ||
MPA (mg/L) | 3 | MPA-1 | 0.76 | 0.06 | 0.06 | 0.07 | 9.3 | 0.77 | 0.04 | 0.07 | 0.07 | 9.6 |
MPA-2 | 1.94 | 0.14 | 0.18 | 0.21 | 10.7 | 1.90 | 0.11 | 0.17 | 0.19 | 9.8 | ||
MPA-3 | 5.16 | 0.08 | 0.12 | 0.14 | 2.6 | 5.10 | 0.06 | 0.04 | 0.06 | 1.1 |
Abbreviations: CSA, cyclosporine; EVE, everolimus; MPA, mycophenolic acid; TAC, tacrolimus; SIR, sirolimus.
Additional survey results for the calibrator reconstitution protocols are shown in Table 3. Based on the survey results, Conditions A0 (similar to Lab A protocol) and B0 (similar to Lab H protocol) were selected. The slopes of the calibration curves did not significantly differ between these two conditions (Table 4).
Additional calibrator reconstitution survey results
Laboratory code | Static incubation time, min | Mix time, min | Mixing by | Use of timer (Static/Mix) |
---|---|---|---|---|
A | 30 | 60 | Roller mixer | Yes/Yes |
B | 15 | 15 | Roller mixer | Yes/Yes |
C | 15 | 5 | Roller mixer | Yes/Yes |
D | 10 | 20 | Roller mixer | Yes/Yes |
E | 0 | 60 | Roller mixer | Yes |
F | 20–30 | 10 | Roller mixer | No/Yes |
G | 0 | 45 | Roller mixer | Yes |
H | 0 | 20 | Roller mixer | Yes |
I | 10–15 | 20–30 | Roller mixer | No/No |
J | 0 | 90 | Shaker | Yes |
Calibrator reconstitution and storage experiment results
Experiment | Drug | Condition | Coefficient | Expected concentration at 6 ng/mL (%diff) | ||
---|---|---|---|---|---|---|
Reconstitution | TAC | A0 | Baseline | 6.000 | (Baseline) | |
B0 | 0.00012 (-0.00034–0.00058) | 0.612 | 5.997 | (-0.049) | ||
SIR | A0 | Baseline | 6.000 | (Baseline) | ||
B0 | -0.00005 (-0.00043–0.00032) | 0.772 | 6.002 | (0.029) | ||
EVE | A0 | Baseline | 6.000 | (Baseline) | ||
B0 | -0.00015 (-0.00036–0.00005) | 0.141 | 6.007 | (0.112) | ||
Storage | TAC | A1 | Baseline | 6.000 | (Baseline) | |
A2 | -0.00037 (-0.00102–0.00028) | 0.257 | 6.009 | (0.143) | ||
A3 | -0.00028 (-0.00093–0.00037) | 0.386 | 6.007 | (0.109) | ||
A4 | -0.00022 (-0.00087–0.00043) | 0.496 | 6.005 | (0.086) | ||
SIR | A1 | Baseline | 6.000 | (Baseline) | ||
A2 | -0.00157 (-0.00239–-0.00076) | < 0.001 | 6.033 | (0.555) | ||
A3 | -0.00091 (-0.00173–-0.00009) | 0.030 | 6.019 | (0.321) | ||
A4 | -0.00061 (-0.00143–0.00020) | 0.138 | 6.013 | (0.217) | ||
EVE | A1 | Baseline | 6.000 | (Baseline) | ||
A2 | -0.00043 (-0.00098–0.00012) | 0.121 | 6.013 | (0.222) | ||
A3 | -0.00117 (-0.00172–-0.00062) | < 0.001 | 6.036 | (0.598) | ||
A4 | -0.00075 (-0.00130–-0.00020) | 0.009 | 6.023 | (0.384) |
*
Abbreviations: CI, confidence interval; EVE, everolimus; TAC, tacrolimus; SIR, sirolimus.
In the experiment on storage conditions, the calibration curve slopes of SIR from Conditions A2 and A3 (
This is the first interlaboratory comparison study of LC-MS/MS for immunosuppressive drug TDM in Korea. The test reproducibility was generally good, and the robust reproducibility CVs were slightly greater than 10% for SIR-1, SIR-2, EVE-1, and EVE-2 with low analyte concentrations. The storage conditions experiment showed that the calibration curve slopes for SIR and EVE changed with prolonged storage of the reconstituted calibrator at 5°C; however, the magnitude of the change was small. Thus, other factors might affect their interlaboratory imprecision.
The method survey results showed that the method details varied across laboratories. For sample preparation, most laboratories used simple protein precipitation methods, and none used additional solid-phase extraction. For internal standards, structural analogs, such as ascomycin, cyclosporine D, and indomethacin, were more commonly used than a stable isotope-labeled internal standard. Although some laboratories used similar sample preparation methods and internal standards and the same commercial calibrator, their test performance could vary depending on the specific protocol followed [17].
In our study, the robust reproducibility CVs were > 10% for the SIR-1, SIR-2, EVE-1, and EVE-2 test samples. Similarly, in the participant summary reports of the CAP international PT programs conducted in 2018, the CVs of SIR and EVE measured by LC-MS/MS were higher than those of the other drugs [8, 18-20]. The reconstitution procedure of the commercial calibrator from Chromsystems is not specified by the manufacturer in detail [21], and the reported times for static incubation and mixing varied across laboratories (Table 3). Considering this, we experimentally determined whether calibrator handling affects the calibration curve and contributes to interlaboratory imprecision. In our experiment, reconstitution with a shorter mixing time did not significantly alter the calibration curve slope, suggesting that the impact of the differences in mixing times is relatively small. However, to prevent possible inadequate mixing or unnecessary timeconsuming procedures, the manufacturer’s recommendation for mixing the calibrator should be specified more clearly. Regarding storage conditions, the manufacturer recommends that the reconstituted calibrator can be stored up to one week at 2–8°C. In our experiment, the calibration curve slopes for SIR and EVE changed with prolonged storage (> one week) of the reconstituted calibrator at 5°C. However, the magnitude of the difference was small, suggesting that a certain extent of prolonged storage after reconstitution might not be a major source of bias.
Although the use of a common calibrator for LC-MS/MS measurement of serum 25-hydroxyvitamin D significantly improved interlaboratory agreement [22], sufficient harmonization among LC-MS/MS methods cannot be achieved by simply using the same commercial calibrator set, as demonstrated for TAC by Levine,
Our study has several limitations: first, we did not assign a reference value to the test samples; therefore, we could not assess laboratory bias. Instead, we aimed to assess test reproducibility in this first interlaboratory comparison in Korea. A refined followup study with traceable assigned values of test samples is needed to assess laboratory bias. Our results could serve as the basis for such a study. In addition, we could not evaluate the homogeneity and the stability of all the test samples owing to the insufficient sample volume. However, we produced the test samples at the same time and in the same manner, and TAC-2 representatively showed acceptable homogeneity. To minimize potential issues with stability, the test samples were shipped in a frozen state and were thawed on the day that the laboratories conducted the tests.
It is challenging to reliably assess interlaboratory comparison results with a small number of participating laboratories [27]. Estimation of location and dispersion for very small datasets can be biased to some degree, even with reasonable outlier rejection and robust statistics. The possibility of high variability in dispersion estimates should be noted. For a balanced summary, we described both estimators using general statistics without outlier removal or robust statistics. As most people in charge of LC-MS/ MS in clinical laboratories are members of the Korean Society of Clinical Chemistry, all of whom participated in this study, the possibility of sampling bias is minimal.
This study provides a snapshot of the current status of immunosuppressive drug TDM by LC-MS/MS in Korea. Although the test reproducibility seems to be good, continuous efforts are required to achieve test standardization and harmonization, especially for SIR and EVE. This study could provide background data for further interlaboratory comparison studies and PT. Unlike general chemistry tests, immunosuppressive drug TDM by LC-MS/MS is conducted as a multiplex test by a small number of laboratories. Further studies should be designed with consid-
eration of these conditions.
The authors thank the staff in the participating laboratories for their cooperation and the sample testing.
Kim HK collected and analyzed the data and drafted the manuscript. Lee SG, Chae H, Song SH, Ahn S, Kim S, and Lee SM analyzed the data and interpreted the results. Park HD, Lee YH, Yun YM, and Lee SY contributed to the conception. SC designed and supervised the study. All authors acquired data from each institution, reviewed the manuscript, and provided critical feedback.
No potential conflicts of interest relevant to this article are reported.
This study was supported by the Clinical Mass Spectrometry Research Committee of the Korean Society of Clinical Chemistry.