Comparison of Three Blood Collection Tubes for 35 Biochemical Analytes: The Becton Dickinson Barricor Tube, Serum Separating Tube, and Plasma Separating Tube
2021; 41(1): 114-119
Ann Lab Med 2021; 41(3): 277-284
Published online May 1, 2021 https://doi.org/10.3343/alm.2021.41.3.277
Copyright © Korean Society for Laboratory Medicine.
Soo Young Moon , M.D.1,2, Han Sol Lee
, M.T.3, Min Soon Park
, M.T.3, In-Suk Kim
, M.D., Ph.D.3,4, and Sun Min Lee, M.D.3
1Department of Laboratory Medicine, 2Biomedical Research Institute, Pusan National University Hospital, Busan, Korea; 3Department of Laboratory Medicine, Pusan National University Yangsan Hospital, Yangsan, Korea; 4Department of Laboratory Medicine, Pusan National University School of Medicine, Yangsan, Korea
Correspondence to: Sun Min Lee, M.D.
Department of Laboratory Medicine, Pusan National University Yangsan Hospital, 20 Geumo-ro, Mulgeum-eup, Yangsan 50612, Korea
Tel: +82-55-360-1878
Fax: +82-55-360-1880
E-mail: yaong97@pnuyh.co.kr
Background: We recently introduced the Barricor (BD, Franklin Lakes, NJ, USA) plasma separation tube, which uses a mechanical separator instead of a gel. We evaluated the effects of using the Barricor tube in a stat (statin) laboratory on the results and turnaround time (TAT) of routine chemical tests. We verified the impact of Barricor tube on reducing TAT and providing results similar to those obtained using serum separator tubes (SSTs).
Methods: We collected venous blood samples from 166 outpatients in Barricor tubes and SSTs and measured 28 routine analytes using an AU5800 instrument (Beckman Coulter, Brea, CA, USA). TAT indexes were compared before and after using Barricor tube.
Results: Mean percent differences were <5%, except for alanine aminotransferase , total CO2, high-density lipoprotein, phosphate, total protein, and direct bilirubin. The median TAT decreased from 45 to 38 minutes, and the rate of a TAT >60 minutes decreased from 7.84% to 2.66%, which was approximately one-third of that for SST. The reduction in TAT was attributable to a decrease in centrifugation time. Incomplete clotting and repeated centrifugation, which occurred frequently when using SST, also decreased after using the Barricor tubes.
Conclusions: The Barricor tube is an alternative to SST for routine chemical tests in institutions aiming to reduce TAT, with clinically allowable differences in test results.
Keywords: Barricor tube, Turnaround time, Routine chemical tests, Serum separation tube
Reducing the turnaround time (TAT) is important in clinical laboratories, especially for emergency and outpatient departments, as it can facilitate faster clinical decision making [1]. Improvements in automation of laboratory instruments and transport technology have reduced TAT, especially for the analysis and post-analysis components. For routine chemistry tests, most laboratories receive serum samples that require clotting. Despite the use of a clot-activation gel, serum separator tubes (SSTs) require approximately 10 minutes each for clotting and centrifugation. To reduce the length of this preanalytical step, the plasma separation tube (PST) and Barricor tube (BD, Franklin Lakes, NJ, USA) were developed. A PST usually contains lithium heparin as an anticoagulant, and numerous PSTs are available from different manufacturers [2]. However, a few studies have shown that the degree of hemolysis is higher in the PST than in the SST [3, 4].
Recently Becton, Dickinson and Company (BD; Franklin Lakes, NJ, USA) introduced the Barricor tube, a PST with a mechanical separation device (stopper). The stopper consists of two parts: an elastomer top that stretches during centrifugation, sealing the stopper to the tube walls, and a high-density plastic base that forms a barrier between the cells and plasma through differential buoyancy. The Barricor tube uses lithium heparin as an anticoagulant; therefore, no clotting step is required. In addition, centrifugation time is shorter for the Barricor tube (3–5 minutes) than for the SST (10 minutes). Therefore, using the Barricor tube could considerably reduce TAT.
The efficacy of the Barricor tube and its ability to reduce TAT compared with the SST have been reported recently [5, 6]. We compared the results of routine chemical tests performed using the SST and the Barricor tube and evaluated the change in TAT and the hemolysis index after switching to the Barricor tube in a one-day visiting department, the third-floor unit. Before Barricor tube introduction, an average of 10,200 SSTs were collected per month: 4,500 in the third-floor phlebotomy unit and 5,700 in the first-floor phlebotomy unit. Of these, 4,300 (95%) from the third-floor unit and 2,000 (35%) from the first-floor unit were requested for stat tests, the results of which are required early. Further, the one-day visiting department receives samples from patients and provides results to doctors before the patients meet the doctors, i.e., in one day. Therefore, we aimed to reduce the TAT of stat tests in the third-floor phlebotomy unit.
Pusan National University Yangsan Hospital at Yangsan, is a national university hospital serving the northern residential region of Busan City, Korea. The hospital has 1,200 beds for inpatients and serves 50,000 outpatients per month. The clinical laboratory is located on the third floor of the main hospital building, and a phlebotomy unit for stat (
The preanalytical process in the third-floor phlebotomy unit is as follows. After the patient registers at the phlebotomy unit, a label containing information and barcode is printed and affixed to a blood tube using a GNT7 automated tube labeling system (Energium, Siheung, Korea). After blood collection, the phlebotomist presses a keyboard button labeled “Blood collection finished” to call the next patient and manually classifies stat test tubes in a basket. Stat test tubes are recommended to be trans-
ferred to the central laboratory every 10 minutes. In the laboratory, a technician receives and registers stat test tubes using a barcode reader. Registered SSTs are centrifuged for 10 minutes in a Kubota 4000 instrument (Kubota, Tokyo, Japan) at 2,270 ×
The study included 166 patients (adults older than 18 years) who visited the stat test phlebotomy unit on the third floor between March 17 and March 24, 2017. All included patients provided written informed consent. The Institutional Review Board of Pusan National University Yangsan Hospital approved this study (No. 05-2016-041). Blood was sampled as per the CLSI guidelines GP41 7th edition, using a 21-G needle and Vacutainer holder (BD Life Science, Seoul, Korea) [8]. From each patient, at least 3 mL of venous blood was collected in a VACUETTE Tube 5 mL serum separator (Greiner Bio-One, Kremsmünster, Austria) and another 3 mL in a Barricor tube. After collection, both tubes were centrifuged and transferred to the routine chemistry unit [9]. The 5 mL SST is a 13 × 100 mm polyethylene terephthalate (PET) tube containing an olefin oligomer barrier gel as a separator and dried microscopic silica particles as clot activators, and the 5.5 mL Barricor tube is a 13 × 100 mm PET tube containing 93.5 IU of dried lithium heparin and a mechanical separator.
The 28 analytes were albumin, alanine alkaline phosphatase, alanine aminotransferase (ALT), amylase, aspartate aminotransferase (AST), blood urea nitrogen, calcium (Ca), chloride (Cl), cholesterol (Chol), creatine kinase (CK), carbon dioxide (CO2), creatinine, C-reactive protein, direct bilirubin (Dbil), γ-glutamyl-transferase, glucose, high-density lipoprotein (HDL), potassium (K), lactate dehydrogenase (LDH), low-density lipoprotein, lipase, magnesium (Mg), phosphate (P), sodium (Na), total bilirubin (Tbil), triglyceride, total protein (Tpro), and uric acid (UA). The analytes were measured using an AU5800 instrument (Beckman Coulter, Brea, CA, USA) with the same lot of dedicated reagents.
TAT was compared between four months each before and after the introduction of the Barricor tube (May to August 2017 vs. September to December 2017). In the third-floor phlebotomy unit, the time point at which the “Blood collection finished” button was pressed was considered the collection time. By contrast, in the first-floor phlebotomy unit, the time point at which the barcode was printed was considered the collection time. The time point at which the technician registered the sample before centrifugation was considered the acceptance time. The time point at which the AU5800 reported the results to the interface software was considered the result time. More than 90% of the results were automatically verified and reported to the laboratory information system (LIS) portal, and this time point was recorded as the report time. We used the AST test to monitor TAT because it is one of the most frequently requested tests. A small proportion of SST samples (9.4%) were for tests other than AST, but AST was considered representative of the 28 analytes.
TAT was defined as the interval between the collection time and the report time. Generally, physicians consider TAT to be the interval from order to report; however, for a one-day patient, a sample is typically ordered during the previous visit (potentially weeks or months before the current visit). Therefore, we decided to calculate the entire TAT from blood collection to reporting. Each month, the TAT of AST tests in the firstand third-floor phlebotomy units was analyzed. The median and 90th percentile TAT for stat tests and the proportion of samples that exceeded the TAT limit (60 minutes) were estimated four months before and after the introduction of the Barricor tube. The TAT of the firstfloor phlebotomy unit was presented as a reference, because no switch to the Barricor tube was made in this unit. We aimed to show that no change other than the introduction of the Barricor tube was made in the analytical and post-analytical phases in our laboratory.
For quality control, our laboratory requires that technicians document the reason for a sample exceeding the TAT limit in the LIS. Beginning in June 2017, 14 reasons were documented (e.g., incomplete clotting and centrifugation, instrument inspection, reagent problem, error flag, delayed result verification, and sample dilution). The number of blood tubes that exceeded the TAT limit and their reasons were retrospectively investigated on the basis of sample and not analyte during the same period.
We measured the hemolysis index semi-quantitatively using the Beckman Coulter LIH reagent in the AU5800 instrument. According to the manufacturer’s instructions, the measured hemolysis index was classified as N, +, ++, and +++, which represent approximately < 0.5, 0.5–0.99, 1–1.99, and 2–2.99 g/L of hemoglobin (a higher degree of hemolysis was not observed in the study period). To estimate hemolytic samples, the ratio of a positive hemolysis index (+, ++, and +++) to the total requested samples was compared before and after introduction of the Barricor tube. Outpatient results with a positive hemolysis index were usually reported with a warning comment. To ensure comparability, only tubes requested for stat tests in the third-floor phlebotomy unit in the two months before and after introduction of the Barricor tube were analyzed.
Bland–Altman difference was analyzed to estimate relative differences between test results of SST and Barricor and 95% limits of agreement for the difference. Because none of the analytes exhibited Gaussian distributions, the Wilcoxon signed-rank test was used to compare medians. The mean difference between the SST and the Barricor tube was compared to allowable bias, which was calculated from the recently updated biological variation provided by the European Federations of Laboratory Medicine (EFLM) [10]. The allowable bias equals the desirable bias, given on the website as desirable bias = 0.25 √(
For the 166 paired clinical samples, most analytes showed acceptable bias between the SST and the Barricor tube (Table 1). The mean percent difference based on the Bland–Altman plot was < 5% for 22 analytes and > 5% for six (ALT, CO2, Dbil, HDL, P, and Tpro). Other electrolytes and proteins were present in a narrow range; therefore, a bias plot was generated to interpret the differences. Compared with those in the SST, Tpro level increased and HDL, K, and P levels decreased in the Barricor tube to within the tolerable range (Fig. 1).
Routine chemical test results using the SST and the Barricor tube (N=166)
Analyte | Unit | Measured range | Mean % difference | 95% limit of agreement | Allowable % bias | Comments on unallowable bias | |
---|---|---|---|---|---|---|---|
Alb | g/L | 24.1–53.5 | −0.19 | −4.26–3.88 | 1.43 | 0.5316 | |
ALP | μkat/L | 0.69–7.75 | −3.37 | −7.98–1.24 | 6.10 | < 0.0001 | |
ALT | μkat/L | 0.14–5.37 | −8.57 | −27.93–10.78 | 7.75 | < 0.0001 | Low ALT level |
Amy | μkat/L | 0.16–3.04 | −0.24 | −5.45–4.96 | 7.73 | 0.3384 | |
AST | μkat/L | 0.16–5.86 | −2.34 | −16.92–12.24 | 5.63 | < 0.0001 | |
BUN | mmol/L | 3.25–45.7 | −0.51 | −3.93–2.9 | 6.15 | 0.0003 | |
Ca | mmol/L | 1.89–2.48 | −1.35 | −5.24–2.54 | 0.82* | < 0.0001 | Strict allowable bias criteria |
Chol | mmol/L | 2.06–9.96 | −2.17 | −5.8–1.46 | 4.52 | < 0.0001 | |
CK | μkat/L | 0.21–9.81 | −0.82 | −14.14–12.5 | 8.90 | 0.0005 | |
Cl | mmol/L | 95.6–112.2 | 0 | −2.17–2.16 | 0.43 | 0.0547 | |
CO2 | mmol/L | 14.8–29 | 18.12 | −1.93–38.18 | 1.68* | < 0.0001 | Impaired stability, strict bias criteria |
Cre | μmol/L | 38.9–746.1 | −2.26 | −8.14–3.61 | 3.75 | < 0.0001 | |
CRP | mg/L | 0.07–189 | −1.85 | −14.29–10.6 | 23.65 | < 0.0001 | |
Dbil | μmol/L | 0.51–226.29 | 13.4 | −78.49–105.3 | 14.19* | 0.053 | |
GGT | μkat/L | 0.14–29.02 | −1.74 | −11.86–8.38 | 10.56 | < 0.0001 | |
Glu | mmol/L | 4.30–17.65 | −1.35 | −5.71–3.01 | 2.38 | < 0.0001 | |
HDL | mmol/L | 0.57–3.07 | −5.06 | −15.19–5.08 | 6.34 | < 0.0001 | Allowable but significant |
K | mmol/L | 3.45–6.42 | −2.69 | −9.91–4.53 | 1.47 | < 0.0001 | Unallowable bias |
LDH | μkat/L | 4.11–38.6 | −0.87 | −15.09–13.36 | 3.38 | 0.0167 | |
LDL | mmol/L | 0.61–251.3 | −1.7 | −6.84–3.45 | 6.85 | < 0.0001 | |
Lip | μkat/L | 0.07–2.43 | −1.29 | −7.96–5.38 | 6.61 | < 0.0001 | |
Mg | mmol/L | 0.51–1.10 | −1.72 | −7.22–3.78 | 1.84* | < 0.0001 | |
Na | mmol/L | 126.7–147.8 | 0.13 | −2.19–2.45 | 0.30 | 0.0001 | |
P | mmol/L | 0.77–1.78 | −4.02 | −9.71–1.67 | 3.38* | < 0.0001 | Unallowable bias |
Tbil | μmol/L | 4.45–470.19 | 1.24 | −5.92–8.4 | 8.95* | 0.0004 | |
TG | mmol/L | 0.41–8.59 | −2.14 | −5.44–1.15 | 10.23 | < 0.0001 | |
Tpro | g/L | 55.3–88.2 | 4.24 | −0.64–9.12 | 1.32 | < 0.0001 | Unallowable bias |
UA | mmol/L | 0.12–0.56 | −0.2 | −3.26–2.87 | 4.87* | 0.7904 |
Mean percent differences (Barricor tube minus SST) and 95% limits of agreement were calculated from Bland–Altman plots.
Abbreviations: SST, serum separator tubes; CI, confidence interval; Alb, albumin; ALP, alanine alkaline phosphatase; ALT, aminotransferase; Amy, amylase; AST, aspartate aminotransferase; BUN, blood urea nitrogen; Ca, calcium, Cl, chloride; Chol, cholesterol; CK, creatine kinase; CO2, carbon dioxide; Cre, creatinine; CRP, c-reactive protein; Dbil, direct bilirubin; GGT, γ-glutamyltransferase; Glu, glucose; HDL, high-density lipoprotein cholesterol; K, potassium; LDH, lactate dehydrogenase; LDL, low-density lipoprotein cholesterol; Lip, lipase; Mg, magnesium; Na, Sodium; P, phosphate; Tbil, total bilirubin; TG, triglyceride; Tpro, total protein; UA, uric acid; EFLM, European Federations of Laboratory Medicine.
We found significant differences between the SST and the Barricor tube for 17 out of the 22 analytes (
In summary, most analytes showed clinically acceptable differences after using the Barricor tube, with a few notable changes.
The number of requested tests, median and 90th percentile TAT, and number and proportion of tests that exceeded the TAT limit (60 minutes) were monitored monthly for eight months in the two phlebotomy units (Table 2). The proportion of tests that exceeded the TAT limit in total tests was regarded the “over rate.” After introducing the Barricor tube in the third-floor phlebotomy unit, the median TAT significantly decreased by 7 minutes, and the 90% TAT decreased by 8 minutes. The major contributor to the reduction in TAT was the reduction in centrifugation time, from 10 minutes to 5 minutes. The over rate decreased from 7.84% to 2.66%, indicating that the number of samples delayed excessively was reduced by one-third. This effect was most noticeable in the first month after introducing the Barricor tube. The number of tests that exceeded the TAT limit decreased by 25% in the first month of Barricor tube use (September), but increased slightly in the next month (October).
TAT after introduction of the Barricor tube (September 2017) in the phlebotomy unit for outpatient stat tests
Third-floor phlebotomy unit (introduced Barricor tube) | First-floor phlebotomy unit (did not introduce Barricor tube) | |||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|
Tests (N) | Median TAT (min) | 90th percentile TAT (min) | Number of times exceeding TAT limit | Over rate (%) | Tests (N) | Median TAT (min) | 90th percentile TAT (min) | Number of times exceeding TAT limit | Over rate (%) | |||
May | 4,052 | 45 | 59 | 327 | 8.07 | 1,538 | 51 | 78 | 420 | 27.31 | ||
June | 4,283 | 45 | 57 | 283 | 6.61 | 1,562 | 52 | 77 | 448 | 28.68 | ||
July | 4,148 | 45 | 59 | 346 | 8.34 | 1,627 | 52 | 73 | 454 | 27.90 | ||
August | 4,249 | 44 | 59 | 355 | 8.35 | 1,730 | 52 | 78 | 513 | 29.65 | ||
September | 4,302 | 36 | 47 | 51 | 1.19 | 1,477 | 52 | 76 | 433 | 29.32 | ||
October | 3,530 | 41 | 55 | 173 | 4.90 | 1,326 | 55 | 77 | 460 | 34.69 | ||
November | 4,154 | 38 | 50 | 95 | 2.29 | 1,479 | 50 | 75 | 380 | 25.69 | ||
December | 4,085 | 39 | 51 | 108 | 2.64 | 1,454 | 54 | 79 | 468 | 32.19 | ||
Before introduction (May–Aug) | 16,732 | 45 | 59 | 1,311 | 7.84 | 6,457 | 52 | 76 | 1,835 | 28.42 | ||
After introduction (Sep–Dec) | 16,071 | 38 | 51 | 427 | 2.66 | < 0.0001 | 5,736 | 53 | 76 | 1,741 | 30.35 | 0.48 |
The TAT was calculated as the time interval from the end of blood collection to the reporting of the initial results. AST was selected to monitor the TAT. Data from the first-floor phlebotomy unit, in which the Barricor tube was not introduced, for the same period are presented as a reference.
Abbreviations: TAT, turnaround time; AST, aspartate aminotransferase.
The TAT of stat tests requested in the first-floor phlebotomy unit was not significantly different in the same period (
The over rate due to insufficient clotting and repeated centrifugation decreased after introduction of the Barricor tube. Incomplete clotting and repeated centrifugation contributed to one-fifth (196/924 = 21%) of the excessive TAT before the introduction of the Barricor tube. However, these reasons for excessive TAT disappeared after the introduction of the Barricor tube (2/210 = 1%). Other reasons, such as delayed sample transport, delayed result verification, and instrument maintenance, also decreased (data not shown).
The hemolysis index decreased significantly after introducing the Barricor tube. In the two months before introducing the Barricor tube, 0.77% (70/9,104) of the samples requested from the third-floor phlebotomy room had a positive hemolysis index (+, ++, or +++), compared with 0.41% (35/8,459) in the two months after its introduction (
We introduced the Barricor tube in an outpatient phlebotomy unit located next to a laboratory in a tertiary care hospital with the aim to reduce TAT. Initially, in September 2017, we used approximately 4,000 Barricor tubes per month; currently, we use nearly 6,000.
The Barricor tube yielded a slightly higher Tpro level and slightly lower HDL, P, and K levels than the SST; however, most values were within the total allowable error (TEA). Tpro level in the Barricor tube was 4.24% higher than that in SST; this could be attributed to the lack of fibrinogen in clotted serum. The K level in the Barricor tube was 2.69% lower than that in the SST, consistent with the fact that serum K level is 0.2–0.4 mmol/L higher than that in plasma due to platelet rupture during clotting. The CO2 value was lower in the SST because CO2 was analyzed after the requested tests were completed and evaporation had occurred.
The Barricor tube has been compared with other tubes [12-18]. A comparison with PST for 50 analytes in 150 patients revealed that the results for the two tubes were comparable, with increased stability for some analytes in the Barricor tube [12]. In 41 volunteers, the Barricor tube yielded decreased LDH and free hemoglobin levels compared with that in the PST [13]. Similar phenomenon of lower LDH in the Barricor tube compared with that in the SST was also observed in the recent Korean study including 30 healthy individuals [14]. In another comparative study using SST in 29 chronic hemodialysis patients, 11 had lower K levels in the Barricor tube; this finding was consistent with our results [15]. A comparison with the SST for 22 analytes in 44 blood samples showed significant increases in LDH and Tpro and a decrease in K in the Barricor tube [16]. Compared with the PST, 40 Barricor tube samples showed a minimal mean difference, except for a 3.2% increase in AST [17]. Another comparative study with the PST on nine analytes in 15 patients showed acceptable differences [18]. Relative to previous studies on the Barricor tube, ours is the largest, with 166 patients. Although there were significant differences in analytes between the two tubes, they were mostly within the clinically acceptable range.
The TAT was reduced by 7 minutes in terms of the median and 8 minutes in terms of the 90th percentile after introducing the Barricor tube. While using SST, incomplete clotting led to repeated centrifugation, fibrin removal, and clogging of the analyzer aspirators. After using the Barricor tube, incomplete clotting and repeated centrifugation were reduced almost to zero. Moreover, the rate of exceeding the TAT limit decreased by onethird, and the medical technicians stated that the repeated centrifugation and analyzer needle clogging decreased after the switch. These improvements can be attributed to the additional 2–3 minutes reduction in TAT besides the 5 minutes decrease in the centrifugation time. If the clotting time had previously been followed as recommended, the reduction in TAT after the switch might have been even more noticeable. However, the real clotting time may be shorter than 30 minutes in most laboratories experiencing TAT pressures.
Some studies have investigated the effects of the Barricor tube on TAT in the phlebotomy unit and clinical laboratory [5, 6]. In a comparison of TAT data during two six-month phases of using the SST and the Barricor tube separately, the Barricor tube significantly reduced the 90th percentile TAT from 88 to 79 minutes, especially during peak hours. The Barricor tube reduced fibrin-related remediation activities from 2.3% to 0.4%, which is consistent with our results [5]. In another study regarding the switch from PST to the Barricor tube, a reduction in centrifugation time from 15 minutes to 3 minutes led to a significant reduction in TAT in an emergency department [6]. In the previous studies, TAT typically began at sample receipt, not blood collection or test request, because this stage is influenced solely by the laboratory [19]. In our study, TAT was measured from the time point at which blood collection was completed, which reflects the real-world situation; this is a strength and a unique aspect of our study.
The plastic mechanical separator in the Barricor tube may promote hemolysis. The Barricor tube has a lower hemolysis rate than other lithium heparin tubes when blood is collected via a Vacutainer, syringe, or intravenous catheter [19-21] because of the low negative pressure in the tube. Indeed, the switch to the Barricor tube resulted in a decreased hemolysis index. This lower hemolysis rate may explain the lower K level than that in the SST.
This study has several limitations. As we did not adhere to the recommended 30 minutes waiting time for clotting when using the SST, the reduction in TAT was smaller than the theoretical expectation. If we had adhered to the 30 minutes waiting time for clotting (including 17 minutes of waiting on average), the TAT would have been reduced by 24 minutes. The TAT and measurement ranges of samples from the intensive care unit and emergency department were not verified. These departments endeavor to decrease TAT and increase satisfaction, which is important for quality improvement [22]. Because our hospital does not accept gel-containing lithium heparin tubes for routine chemical tests, we did not perform a comparison with the PST. The Barricor tube reduces the rate of false elevation of troponin I compared with the PST and does not induce isobaric interference [23, 24]. We introduced the Barricor tube to reduce the TAT in the outpatient phlebotomy unit for routine chemical tests; thus, further investigations of other uses of the Barricor tube are needed. A stability test was not performed, as we applied the Barricor tube for stat tests, which took approximately 45 minutes from collection to reporting.
In conclusion, the Barricor tube showed allowable differences compared with the SST. After introducing the Barricor tube, TAT was reduced owing to decreased centrifugation time and clogging-related problems. The Barricor tube might be useful in laboratories that experience frequent clogging and repeated centrifugation due to inadequate SST clotting time. Further investigation is needed on the cost-effectiveness and other unrevealed applications of the Barricor tube to reduce TAT.
We thank Kinam Park of the Information Technology Department of Pusan National University Yangsan Hospital for extracting LIS data. In addition, we thank Seung Yeon Lee and the staff at BD Life Science, Korea, for providing Barricor tubes for evaluation and for technical assistance.
This study was designed and organized by Lee SM. Moon SY wrote the first draft and performed the statistical analysis. Lee HS and Park MS collected comparative results and turnaround time data. Kim IS and Lee SM provided technical support for the study as directors of the clinical laboratory and stat laboratory, respectively. All authors reviewed and approved the final manuscript.
Barricor tubes were provided by BD Life Science, Korea. No further potential conflicts of interest relevant to this article are reported.
This work was supported by a clinical research grant from Pusan National University Hospital (2019).